FDAFDA · Food and drugs·Recall number Z-0003-2015

Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Published October 8, 2014 · Recall initiated: September 2, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
Lot
Siemens AXIOM Luminos dRF Max system material #s: 10762471 with serial number: 5037.
Risk classification
Class II