SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. For intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.
Published October 7, 2015 · Recall initiated: September 10, 2015
Summary
FDA records medical device Class II recall Z-0003-2016, published October 7, 2015 and initiated September 10, 2015. Recalling firm: Biosense Webster, Inc.; Status: Terminated; Reported quantity: 1035 (U.S.); Lot: Catalog #: 10439236, 10439072, 10439011, 10438577. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biosense Webster, Inc.
- Distributor
- Worldwide Distribution. US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.
- Lot
- Catalog #: 10439236, 10439072, 10439011, 10438577
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheath isMarch 20, 2024
- FDABiosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CEMarch 31, 2021
- FDAWaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.September 5, 2018
- FDASMARTABLATE RF System GeneratorDecember 13, 2017
- FDAImperial GMI, Non GMO Linerless Sugar EFG, Net Weight 2000 lbs., Imperial Sugar Company, Savannah, GA.October 7, 2015
- FDAZimmer, Reprocessed Zimmer Tourniquet Cuffs,34" Single Hose Single Bladder Brown (PLC), 34", ModelOctober 7, 2015
Questions about this recall
Was SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. For intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. recalled?
Yes. FDA published recall Z-0003-2016 on October 7, 2015. Reported hazard: Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
What should I do if I have this product?
Voluntary: Firm initiated · Letter