Back to search
FDAFDA · Food and drugs·Recall number Z-0003-2016

SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. For intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.

Published October 7, 2015 · Recall initiated: September 10, 2015

Summary

FDA records medical device Class II recall Z-0003-2016, published October 7, 2015 and initiated September 10, 2015. Recalling firm: Biosense Webster, Inc.; Status: Terminated; Reported quantity: 1035 (U.S.); Lot: Catalog #: 10439236, 10439072, 10439011, 10438577. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biosense Webster, Inc.
Distributor
Worldwide Distribution. US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.
Lot
Catalog #: 10439236, 10439072, 10439011, 10438577
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. For intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. recalled?

Yes. FDA published recall Z-0003-2016 on October 7, 2015. Reported hazard: Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.

What should I do if I have this product?

Voluntary: Firm initiated · Letter