Back to search
FDAFDA · Food and drugs·Recall number Z-0004-2015

Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: "Querying and retrieving patient history information and/or previous diagnosis and images from other modalities; "X-ray examinations of the urogenital area "Ultrasound examinations "Endourological interventions "Percutaneous interventions "Laparoscopy "Application of fistula "Simple procedures "Extracorporeal shock wave lithotripsy "Uroflow/urodynamics "Pediatric radiological and therapeutic applications

Published October 8, 2014 · Recall initiated: September 2, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
Lot
Siemens Uroskop Omnia Max system, material #s: 10762473, serial numbers: 4014 4015
Risk classification
Class II