Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in endoscopic procedures of the upper and lower gastrointestinal tract or biliary tract.
Published October 10, 2012 · Recall initiated: November 14, 2011
Summary
FDA records medical device Class II recall Z-0006-2013, published October 10, 2012 and initiated November 14, 2011. Recalling firm: US Endoscopy Group Inc; Status: Terminated; Reported quantity: 1153; Lot: Model #00711177; S/N: 52979, 52981, 52982, 53643, 53644, 53891, 53892, 54238. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- US Endoscopy Group Inc
- Distributor
- Worldwide Distribution -- USA (nationwide) including the states of AL, AR, AZ, CA, FL, GA, HI, IA, IL, KY, LA, MD, ME, MI, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, and WI; and to the following countries: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Caribbean, Chile, Columbia, Croatia, Czech Republic, Denmark, Ecuador, El Salvador, Estonia, Finland, France, Germany, Great Britain, Greece, India, Ireland, Israel, Italy, Jordan, Lithuania, Malta, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, and United Arab Emirates.
- Lot
- Model #00711177; S/N: 52979, 52981, 52982, 53643, 53644, 53891, 53892, 54238
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVari-Safe Injection Needle US endoscopy 5976 Heisley Road Mentor, Ohio 44060 A sterile, single use, flexibleOctober 26, 2016
- FDAUroSeal Adjustable Endoscopic Valve. Model/Catalog # 00913410 The primary packaging is a Tyvek pouch. TheDecember 31, 2014
- FDATorrent" irrigation tubing. Product Usage: The Torrent irrigation system (tubing and accessories toApril 17, 2013
- FDAIR1200, IR1250(IR1200 Plus), 2020 and One Touch(R) Ping(R) Insulin Pump glucose management system. The deviceOctober 10, 2012
- FDABriteField illuminated Suction Device; Model numbers 8Fr, 10Fr and 12Fr, Kit (included 8Fr, 10Fr and 12Fr andOctober 10, 2012
- FDAInfinity® sampling device, Cytology Brush, United States Endoscopy Group, Inc., 5976 Heisley Road, Mentor, OHSeptember 26, 2012
Questions about this recall
Was Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in endoscopic procedures of the upper and lower gastrointestinal tract or biliary tract. recalled?
Yes. FDA published recall Z-0006-2013 on October 10, 2012. Reported hazard: When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition.
What should I do if I have this product?
Voluntary: Firm initiated · Letter