Sterile, single-use, disposable electrosurgical electrode holder. It connects the electrode to the foot activated electric generator. Label reads in part "***GYNEX*** Electrode Handpiece, Foot Activated*** Manufactured for GYNEX***" Gynex catalog number: 950 (Gynex REF 950, Electrode Handpiece) It connects the electrode to the food activated electric generator. Electrodes are used in gynecological procedures, in a physician's office, for cutting to remove tissue and coagulation to control bleeding by the use of high frequency current.
Published October 10, 2012 · Recall initiated: September 6, 2012
Summary
FDA records medical device Class III recall Z-0008-2013, published October 10, 2012 and initiated September 6, 2012. Recalling firm: Gynex Corporation; Status: Terminated; Reported quantity: 19 boxes (containing 12 electorde hand pieces each); Lot: Manufacturer's lot number: 54-12-2-21-12. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The wrong electrical connector pin was attached to the cord during assembly by the manufacturer. this electrical connector is too big to fit into the port on the foot activated generator. this makes the device unusable.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Gynex Corporation
- Distributor
- Nationwide Distribution
- Lot
- Manufacturer's lot number: 54-12-2-21-12.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABriteField illuminated Suction Device; Model numbers 8Fr, 10Fr and 12Fr, Kit (included 8Fr, 10Fr and 12Fr andOctober 10, 2012
- FDAWound Pro Apex Negative Pressure Wound Therapy Powered Suction Pump; Accuro Medical Products LLC, 2854October 10, 2012
- FDAEndoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies orOctober 10, 2012
- FDA8 Fr. Bardex Pediatric All-Silicone Foley Catheter with a 3cc balloon. The product is packaged individuallyOctober 10, 2012
- FDAIR1200, IR1250(IR1200 Plus), 2020 and One Touch(R) Ping(R) Insulin Pump glucose management system. The deviceOctober 10, 2012
- FDAPhilips HeartStart XL M4735A Defibrillator/Monitors, used with the M3516A batteries The Heart Start XLOctober 10, 2012
Questions about this recall
Was Sterile, single-use, disposable electrosurgical electrode holder. It connects the electrode to the foot activated electric generator. Label reads in part "***GYNEX*** Electrode Handpiece, Foot Activated*** Manufactured for GYNEX***" Gynex catalog number: 950 (Gynex REF 950, Electrode Handpiece) It connects the electrode to the food activated electric generator. Electrodes are used in gynecological procedures, in a physician's office, for cutting to remove tissue and coagulation to control bleeding by the use of high frequency current. recalled?
Yes. FDA published recall Z-0008-2013 on October 10, 2012. Reported hazard: The wrong electrical connector pin was attached to the cord during assembly by the manufacturer. this electrical connector is too big to fit into the port on the foot activated generator. this makes the device unusable.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit