Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Published October 8, 2025 · Recall initiated: September 3, 2025
Summary
FDA records medical device Class II recall Z-0009-2026, published October 8, 2025 and initiated September 3, 2025. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 87 units (21 US, 66 OUS); Lot: Model Numbers: 722029; UDI-DI: 00884838054219; Serial Numbers: 71, 91, 7, 94, 62, 88, 36, 97, 21, 100, 30, 63, 60, 129, 95, 126, 110, 56, 55, 57, 31, 26, 49, 114, 128, 108, 103, 107, 81, 43, 52, 119, 118, 80, 78, 47, 106, 123, 96, 122, 14, 127, 101, 116, 102, 115, 19, 109, 120, 92, 69, 27, 46, 93, 51, 50, 124, 87, 20, 29, 32, 104, 34, 45, 59, 79, 24, 70, 39, 66, 82, 75, 40, 33, 23, 25, 85, 37, 44, 58, 38, 61, 98, 53, 89, 35, 48. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T®rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
- Lot
- Model Numbers: 722029; UDI-DI: 00884838054219; Serial Numbers: 71, 91, 7, 94, 62, 88, 36, 97, 21, 100, 30, 63, 60, 129, 95, 126, 110, 56, 55, 57, 31, 26, 49, 114, 128, 108, 103, 107, 81, 43, 52, 119, 118, 80, 78, 47, 106, 123, 96, 122, 14, 127, 101, 116, 102, 115, 19, 109, 120, 92, 69, 27, 46, 93, 51, 50, 124, 87, 20, 29, 32, 104, 34, 45, 59, 79, 24, 70, 39, 66, 82, 75, 40, 33, 23, 25, 85, 37, 44, 58, 38, 61, 98, 53, 89, 35, 48;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAPhilips Azurion 5 M12. Model Number: 722227.November 4, 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.November 4, 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.November 4, 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.November 4, 2026
- FDAAllura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638,October 8, 2025
- FDAAllura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979,October 8, 2025
Questions about this recall
Was Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563; recalled?
Yes. FDA published recall Z-0009-2026 on October 8, 2025. Reported hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
What should I do if I have this product?
Voluntary: Firm initiated · Letter