Philips HeartStart XL M4735A Defibrillator/Monitors, used with the M3516A batteries The Heart Start XL Defibrillator/Monitor is a fully featured, defibrillator intended for use by qualified medical personnel, trained in either Advanced Cardiac Life Support or in the operation of the device, in a hospital environment.
Published October 10, 2012 · Recall initiated: September 11, 2012
Summary
FDA records medical device Class II recall Z-0011-2013, published October 10, 2012 and initiated September 11, 2012. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 2500 units; Lot: HeartStart XL battery M3516A labeled with Made in Taiwan and an R-2011-12 Date of Manufacture. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
HeartStart XL Defibrillator/Monitor (M4735A) used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare Inc.
- Distributor
- Worldwide distribution: USA (Nationwide) and countries of: AUSTRALIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, REPUBLIC OF KUWAIT LEBANON, MACAU, MALAYSIA, MEXICO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, ROMANIA RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, and VIET NAM.
- Lot
- HeartStart XL battery M3516A labeled with Made in Taiwan and an R-2011-12 Date of Manufacture
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMobile Diagnostic X-Ray System with Digital PanelApril 12, 2017
- FDAPhilips IntellVue Info Center iX 866023; PIIC iX Upgrade 866024; IntellVue Server iX Upgrade 866026; PIICJune 18, 2014
- FDAPhilips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.September 4, 2013
- FDAPhilips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage:August 28, 2013
- FDA8 Fr. Bardex Pediatric All-Silicone Foley Catheter with a 3cc balloon. The product is packaged individuallyOctober 10, 2012
- FDAEOS X-ray Medical Diagnostic , X-ray Equipment System Digital Radiography used in general radiographicOctober 10, 2012
Questions about this recall
Was Philips HeartStart XL M4735A Defibrillator/Monitors, used with the M3516A batteries The Heart Start XL Defibrillator/Monitor is a fully featured, defibrillator intended for use by qualified medical personnel, trained in either Advanced Cardiac Life Support or in the operation of the device, in a hospital environment. recalled?
Yes. FDA published recall Z-0011-2013 on October 10, 2012. Reported hazard: HeartStart XL Defibrillator/Monitor (M4735A) used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.
What should I do if I have this product?
Voluntary: Firm initiated · Letter