FDAFDA · Food and drugs·Recall number Z-0011-2014

ADVIA 2120/2120i Hematology Systems with ADVIA Autoslide connected to it, SMN 10361162 and SMN 10361798 --- CLASSIFICATION NAME: Counter, Differential Cell, Automated (Particle Counter) The ADVIA 2120/2120i are hematology systems that utilizes the principles of flow cytometry in order to provide complete blood counts.

Published October 23, 2013 · Recall initiated: August 12, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Siemens determined that if an autosampler rack jam error occurs during operation of an ADVIA 2120/2120i connected to an ADVIA Autoslide, it is possible that the next slide processed by the Autoslide could be labeled with the wrong sample identification information.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Healthcare Diagnostics
Distributor
Worldwide Distribution - USA (nationwide) and the countries of Australia, Austria, Belgium, Canada, Croatia, Denmark, France, Germany, Greece, Italy, Japan, Malaysia, Netherlands, Norway, Philippines, Portugal, South Korea, Singapore, Slovenia, Spain, Switzerland, Thailand, Turkey, United Kingdom and Vietnam.
Lot
All serial numbers connected to the ADVIA 2120/2120i Hematology Systems with ADVIA Autoslide.
Risk classification
Class II
Area (FDA)
device
Country
United States