Back to search
FDAFDA · Food and drugs·Recall number Z-0013-2017

Multix Fusion Stationary X-Ray System Product Usage: The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill)applications.

Published October 12, 2016 · Recall initiated: August 1, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Siemens became aware of a potential issue with the patient table for Ysio Max and Multix Fusion systems. There is an unlikely risk that an internal part of the table may detach causing the table top to tilt to the left side which could result in an injury to a patient on the table.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distribution
Lot
Model Number: 10762470, 10893300, 10746665 Serial Numbers: 24644,24613,24615,40328,24611,24612,24614,40326,40322,40316,24599,24600,40404,24634,2464124590,40402,1314,40312,24591,24592,40327,40323,24656,24659,40335,1315,40332,40403,24605,40325,40313,24645,24610,24580,24648,24581
Risk classification
Class II
Area (FDA)
device
Country
United States