FDAFDA · Food and drugs·Recall number Z-0014-2014
Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle.
Published October 23, 2013 · Recall initiated: March 8, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A post market safety review following several customer complaints indicated additional clarification was necessary in the Midfoot Fusion Bolt Technique Guide to reduce potential for bolt back-out or migration by emphasizing the need for supplemental fixation.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Nationwide Distribution, including the states of MI, PA, SC, IA, MA, CA, WI, IL, and KS.
- Lot
- Part Nos.: 04.111.210, 04.111.220, 04.111.230. with Lot Nos.; 2294209 through 7587547.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States