FDAFDA · Food and drugs·Recall number Z-0016-2013

Active Renin IRMA Kit, Part Number DSL25100 Product Usage: The Renin IRMA; DSL25100 immunoradiometric assay kit provides materials for the quantitative measurement of Renin in plasma. This assay is intended for in vitro diagnostic use.

Published October 17, 2012 · Recall initiated: January 20, 2012

Summary

FDA records medical device Class II recall Z-0016-2013, published October 17, 2012 and initiated January 20, 2012. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 216 units total (6 units in the US); Lot: Lot # 110919D, 111017C, 111017D, 111114C, 111212C. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The recall was initiated because Beckman Coulter has confirmed that the Active Renin IRMA Test Kit may exhibit a decrease in stability of the coated tubes. As a result, quality control and patient sample results may be falsely elevated by up to 72%.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Worldwide Distribution - US Natiowide including the state of: OH, CT, CA and the state of Argentina, Austria, Belgium, Brazil, Canada, China, Columbia, Czech Republic, France, Germany, Greece, Hungary, Italy, Kuwait, Netherlands, Poland, Portugal, Slovakia, Spain, Switzerland, Turkey, and United Kingdom.
Lot
Lot # 110919D, 111017C, 111017D, 111114C, 111212C.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Active Renin IRMA Kit, Part Number DSL25100 Product Usage: The Renin IRMA; DSL25100 immunoradiometric assay kit provides materials for the quantitative measurement of Renin in plasma. This assay is intended for in vitro diagnostic use. recalled?

Yes. FDA published recall Z-0016-2013 on October 17, 2012. Reported hazard: The recall was initiated because Beckman Coulter has confirmed that the Active Renin IRMA Test Kit may exhibit a decrease in stability of the coated tubes. As a result, quality control and patient sample results may be falsely elevated by up to 72%.

What should I do if I have this product?

Voluntary: Firm initiated · Letter