The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode leads. It is powered by a hermetically sealed battery within a titanium case. It uses microelectronic circuitry to generate constant current electrical stimulation. Stimulation programs can be delivered as either single stimulation or MultiStim programs depending on patients need. Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
Published October 17, 2012 · Recall initiated: September 4, 2012
Summary
FDA records medical device Class II recall Z-0025-2013, published October 17, 2012 and initiated September 4, 2012. Recalling firm: Advanced Neuromodulation Systems Inc.; Status: Terminated; Reported quantity: 56 units; Lot: Eon Mini - 3788 (IPG), Multiple Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Advanced Neuromodulation Systems Inc.
- Distributor
- US Nationwide Distribution
- Lot
- Eon Mini - 3788 (IPG), Multiple Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAdvanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788March 6, 2013
- FDAPREVI Isola System, Catalog numbers 29500 and 29500R, bioMerieux, Inc., Durham, NC. Intended to perform theOctober 17, 2012
- FDAOrthoHelix Part #HCS-056-32 3.2MM Cannulated Hex Driver NON STERILE MAT: 465 SS The drivers are used to driveOctober 17, 2012
- FDAThe Eon Mini Neurostimulation (IPG) System (Model 3788) is a 16-channel, rechargeable, multi-program systemSeptember 5, 2012
- FDAEon (Model 3716) IPG is a dual channel; 8-contact receiver designed to be connected to 4, 8, or 16 electrodeSeptember 5, 2012
- FDAEon Mini Neurostimulation (IPG) System (Model 3788); The product is indicated as an aid in the management ofJuly 18, 2012
Questions about this recall
Was The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode leads. It is powered by a hermetically sealed battery within a titanium case. It uses microelectronic circuitry to generate constant current electrical stimulation. Stimulation programs can be delivered as either single stimulation or MultiStim programs depending on patients need. Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain. recalled?
Yes. FDA published recall Z-0025-2013 on October 17, 2012. Reported hazard: Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.
What should I do if I have this product?
Voluntary: Firm initiated · Letter