Back to search
FDAFDA · Food and drugs·Recall number Z-0028-2026

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Published October 15, 2025 · Recall initiated: September 9, 2025

Summary

FDA records medical device Class I recall Z-0028-2026, published October 15, 2025 and initiated September 9, 2025. Recalling firm: NOXBOX LTD; Status: Ongoing; Reported quantity: 1667 units; Lot: Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
NOXBOX LTD
Distributor
Domestic: TN
Lot
Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.
Risk classification
Class I
Area (FDA)
device
Country
United Kingdom

Related recalls

Questions about this recall

Was NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I recalled?

Yes. FDA published recall Z-0028-2026 on October 15, 2025. Reported hazard: Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail