FDAFDA · Food and drugs·Recall number Z-0030-2013
Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL) Product Usage: Enzyme n Immunoassay, Phenobarbital
Published October 17, 2012 · Recall initiated: July 30, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm has confirmed highter imprecision with the recalled lots on Beckman Coulter AU(R) Systems. Also, if the Syva(R) Emit(R) 2000 Phenobartbital Assay (4D019UL) lots D4, D5, D6, and/or D7 as an application on AU(R) Systems is used, the issue may be experienced. Firm has found that within-run imprecison of the recalled lots when run on an AU(R) System may be higher that what would normally be obs
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- United States Nationwide Distribution in the states of: CA, FL, KS, MA, MN, MS, NM, NY, NC, OR, PA, TN, TX, UT, VA, and WA.
- Lot
- Emit(R) 2000 Phenobarbital Assay (OSR4D229), lots 1151, 1167, 1191, and/or 1212 and Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots D4, D5, D6, and D7
- Risk classification
- Class II