L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.
Published October 9, 2019 · Recall initiated: October 10, 2018
Summary
FDA records medical device Class II recall Z-0030-2020, published October 9, 2019 and initiated October 10, 2018. Recalling firm: Baxter Healthcare Corporation; Status: Terminated; Reported quantity: 8640 eaches; Lot: Lot Number H18D22030. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.
- Lot
- Lot Number H18D22030
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATruSystem 7500 surgical table, Model/Catalog Number: 4091000November 11, 2026
- FDATruSystem 7000 U Operating Table (dV), Product Code 1723633November 11, 2026
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAOperating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788November 11, 2026
- FDAGrandio SO Heavy Flow Caps, Color A3 - Product Usage: Filling minimally invasive cavities of all classes -October 9, 2019
- FDATxRx 15Ch Knee Coil 3T QED part number:Q7000008 Siemens Model # 10185460October 9, 2019
Questions about this recall
Was L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide. recalled?
Yes. FDA published recall Z-0030-2020 on October 9, 2019. Reported hazard: Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.
What should I do if I have this product?
Voluntary: Firm initiated · Letter