Vascular Solutions Twin-Pass .023" Dual Access Catheter, model 5230. Sterilized with Ethylene Oxide. The Twin-Pass dual access catheter is a dual lumen catheter designed for use in the coronary and/or peripheral vasculature. The Twin-Pass catheter (Models 5200 and 5230) consists of an over-the-wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal segment. The Twin-Pass catheter has a working length of 135cm and is compatible with various guidewire diameters. Model 5200 also has hydrophilic coating on the distal 18cm of the catheter. The Twin-Pass catheter has white positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, respectively. The Twin-Pass catheter has a marker band located 1mm from the distal tip. Model 5200 contains a second marker band located 11mm from the distal tip to identify the distal end of the OTW lumen.

Published October 26, 2016 · Recall initiated: September 16, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Vascular Solutions is recalling Twin-Pass dual access catheters because there is a potential for excess manufacturing material to remain at the tip of the catheter or within the distal portion of the rapid exchange lumen.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Vascular Solutions, Inc.
Distributor
Worldwide Distribution: US (nationwide) including Puerto Rico and outside the US to include: Australia, Austria, Belgium, Brazil, Canada, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
Lot
Model 5230, Lot numbers: 577279, 578997, 580613, 583021, 584463, 586310, 588542, 590739, 591262, 593696, 593985.
Risk classification
Class I
Area (FDA)
device
Country
United States