DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.
Published October 14, 2020 · Recall initiated: August 3, 2020
Summary
FDA records medical device Class II recall Z-0060-2021, published October 14, 2020 and initiated August 3, 2020. Recalling firm: Encore Medical, LP; Status: Ongoing; Reported quantity: 190 units; Lot: Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10 Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Encore Medical, LP
- Distributor
- U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.
- Lot
- Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10 Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 342-19-703July 15, 2026
- FDABrand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 341-14-711July 15, 2026
- FDABrand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEE INS, 7R 16MM, VE Model/CatalogAugust 20, 2025
- FDABrand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEETM INS, 11L 12MM, VEJuly 23, 2025
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x andOctober 14, 2020
- FDAKit Model #65021652, MTO Left Heart Kit United Reg Health - Product Usage: NAMIC Cath Lab and InterventionalOctober 14, 2020
Questions about this recall
Was DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel. recalled?
Yes. FDA published recall Z-0060-2021 on October 14, 2020. Reported hazard: There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail