Back to search
FDAFDA · Food and drugs·Recall number Z-0060-2021

DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.

Published October 14, 2020 · Recall initiated: August 3, 2020

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Encore Medical, LP
Distributor
U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.
Lot
Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10 Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card