1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not limited to dental, ENT (ear, nose and throat), neuro, spine and endoscopic applications Catalog Number: 5820-066-012
Published October 23, 2024 · Recall initiated: September 5, 2024
Summary
FDA records medical device Class II recall Z-0063-2025, published October 23, 2024 and initiated September 5, 2024. Recalling firm: Stryker Corporation; Status: Ongoing; Reported quantity: 307 units; Lot: GTIN: 04546540362346 Lot number: 22329017. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Corporation
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, Columbia.
- Lot
- GTIN: 04546540362346 Lot number: 22329017
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASTRYKER PKG., STRYKEFLOW SUCTION/IRRIGATOR PS, Catalog Number 0250070540August 12, 2026
- FDASTRYKER AHTO DUAL SPIKE TUBE SET PACKAGING, Catalog Number 0250070640August 12, 2026
- FDASTRYKER AHTO TUBE SET WITH TIP PACKAGING, Catalog Number 0250070620August 12, 2026
- FDASTRYKER AHTO TUBE SET PACKAGING, Catalog Number 0250070600August 12, 2026
- FDAMelrose Wakefield Hospital REF K05T-2159 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - TheOctober 23, 2024
- FDABlue Diamond Digital Inflation Device, REF: IN7152/C, IN7352/C, IN7403/C, IN7802/C, IN7852/C, IN7130/COctober 23, 2024
Questions about this recall
Was 1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not limited to dental, ENT (ear, nose and throat), neuro, spine and endoscopic applications Catalog Number: 5820-066-012 recalled?
Yes. FDA published recall Z-0063-2025 on October 23, 2024. Reported hazard: May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
What should I do if I have this product?
Voluntary: Firm initiated · Letter