Delta Cup Beater Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System
Published November 15, 2017 · Recall initiated: September 8, 2017
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Limacorporate S.p.A
- Distributor
- Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
- Lot
- Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J
- Risk classification
- Class II