Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System

Published November 15, 2017 · Recall initiated: September 8, 2017

Summary

FDA records medical device Class II recall Z-0068-2018, published November 15, 2017 and initiated September 8, 2017. Recalling firm: Limacorporate S.p.A; Status: Completed; Reported quantity: 20 units in (US); Lot: Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Limacorporate S.p.A
Distributor
Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
Lot
Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J
Risk classification
Class II
Area (FDA)
device
Country
Italy

Related recalls

Questions about this recall

Was Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System recalled?

Yes. FDA published recall Z-0068-2018 on November 15, 2017. Reported hazard: An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

What should I do if I have this product?

Voluntary: Firm initiated · Letter