Back to search
FDAFDA · Food and drugs·Recall number Z-0068-2018

Delta Cup Beater  Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System

Published November 15, 2017 · Recall initiated: September 8, 2017

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Limacorporate S.p.A
Distributor
Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
Lot
Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J
Risk classification
Class II
Area (FDA)
device
Country
Italy

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card