Delta Cup Beater Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System
Published November 15, 2017 · Recall initiated: September 8, 2017
Summary
FDA records medical device Class II recall Z-0068-2018, published November 15, 2017 and initiated September 8, 2017. Recalling firm: Limacorporate S.p.A; Status: Completed; Reported quantity: 20 units in (US); Lot: Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Limacorporate S.p.A
- Distributor
- Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
- Lot
- Lot # 15AG0HJ, 15AG0HQ, 15AG0UY, 15AG00P, 15AG040, 15AG08Y, 14AG243, 15AG00Q, 15AG0UZ, 15AG0V0, 15AG0HN, 16AG02J
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy
Related recalls
- FDAPRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MMJanuary 24, 2024
- FDACobalt HV with Gentamicin, Cobalt Bone Cement 40GM, REF 402283, QTY 1, STERILE, Rx Only, Mfgr: BIOMETNovember 15, 2017
- FDAAERO IOL Injection System containing 10 Z28 Cartridges per box, Part #003500-0025-282, Sterile, Rx. The firmNovember 15, 2017
- FDACobalt MV with Gentamicin, Cobalt Bone Cement 40GM, REF 402439, QTY 1, STERILE, Rx Only, Mfgr: BIOMETNovember 15, 2017
- FDASMR glenosphere impactor/extractor; Used to implant the glenosphere in the SMR Reverse Shoulder replacement.June 1, 2016
- FDASMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mmSeptember 9, 2015
Questions about this recall
Was Delta Cup Beater Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta Acetabular System recalled?
Yes. FDA published recall Z-0068-2018 on November 15, 2017. Reported hazard: An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.
What should I do if I have this product?
Voluntary: Firm initiated · Letter