FDAFDA · Food and drugs·Recall number Z-0071-2013
Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by strategically placed mounting arms. Additionally, there are three port supports the user incorporates to position the lines that enter and exit the venous bag. The Backplate/Holder is designed to allow the user to move the port support location to accommodate different tubing circuit designs to permit mounting flexibility.
Published October 24, 2012 · Recall initiated: October 2, 2009
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one ¿ inch connector supports instead of one 3/8 inch and two ¿ inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a
What to do
Voluntary: Firm initiated · Telephone
Recall details
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Distributor
- Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.
- Lot
- No lot/serial number; Catalog number: PN 816280
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States