SIEMENS IMMULITE/IMMULITE 2000 Systems Rubella IgM (RUM), REF/Catalog Number for US: L2KRM2(D), Siemens Material Number (SMN) for US: 10374021, REF/Catalog # for OUS: L2KRM2, SMN for OUS: 10381327; Unique Device Identification Number (UDI) 00630414961996; IVD --- For in vitro diagnostic use with the IMMULITE/IMMULITE 1000/IMMULITE 2000/IMMULITE 2000 XPi Systems Analyzers-for the qualitative detection of IgM antibodies to Rubella virus in human serum or plasma (EDTA and heparinized), as an aid in the presumptive diagnosis of an acute or recent rubella infection, particularly in women of childbearing age
Published October 19, 2016 · Recall initiated: August 24, 2016
Summary
FDA records medical device Class II recall Z-0082-2017, published October 19, 2016 and initiated August 24, 2016. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: Domestic: 177 Kits; Foreign: 3435 Kits; Lot: Kit lots 236M, 237, 238, 239, 239L (OUS) and D236M and D238 (US). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution n- US (Nationwide) and Foreign distribution to the following countries: Argentina, Bangladesh, Brazil, Bulgaria, China, Colombia, Costa Rica, Cyprus, Egypt, France, Germany, Greece, Guatemala, India, Indonesia, Israel, Italy, Kazakhstan, Macedonia, Mexico, Paraguay, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Africa, Syria, United Arab Emirates, and Uruguay.
- Lot
- Kit lots 236M, 237, 238, 239, 239L (OUS) and D236M and D238 (US)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDATiger-X, 1200 mg, 1 capsule per pack , 24 single packs per boxOctober 26, 2016
- FDATheraSphere(R) YTTRIUM-90 CLASS MICROSPHERE INJECTION TheraSphere® is indicated for radiation treatment or asOctober 19, 2016
Questions about this recall
Was SIEMENS IMMULITE/IMMULITE 2000 Systems Rubella IgM (RUM), REF/Catalog Number for US: L2KRM2(D), Siemens Material Number (SMN) for US: 10374021, REF/Catalog # for OUS: L2KRM2, SMN for OUS: 10381327; Unique Device Identification Number (UDI) 00630414961996; IVD --- For in vitro diagnostic use with the IMMULITE/IMMULITE 1000/IMMULITE 2000/IMMULITE 2000 XPi Systems Analyzers-for the qualitative detection of IgM antibodies to Rubella virus in human serum or plasma (EDTA and heparinized), as an aid in the presumptive diagnosis of an acute or recent rubella infection, particularly in women of childbearing age recalled?
Yes. FDA published recall Z-0082-2017 on October 19, 2016. Reported hazard: Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.
What should I do if I have this product?
Voluntary: Firm initiated · Letter