DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.
Published November 6, 2013 · Recall initiated: July 26, 2013
Summary
FDA records medical device Class II recall Z-0090-2014, published November 6, 2013 and initiated July 26, 2013. Recalling firm: Carestream Health, Inc.; Status: Terminated; Reported quantity: 23 units installed total (11 domestically & 12 internationally); Lot: Serial numbers: U.S. - 134,152,173,174,200,205,222,312,349,353,412, Worldwide Serial numbers: 469, 445, 306, 410, 453, 204, 483, 300, 171, 218. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Carestream Health, Inc.
- Distributor
- Worldwide Distribution - US (nationwide) including the states of New York, Massachusetts, Pennsylvania, Illinois, Texas, California and Louisiana., and the countries of Australia, Canada, Finland, Italy, Switzerland and United Kingdom.
- Lot
- Serial numbers: U.S. - 134,152,173,174,200,205,222,312,349,353,412, Worldwide Serial numbers: 469, 445, 306, 410, 453, 204, 483, 300, 171, 218
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADRX Revolution Mobile X-Ray SystemDecember 13, 2023
- FDAOnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console)May 1, 2019
- FDACARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening andDecember 24, 2014
- FDAKodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently installedAugust 20, 2014
- FDABrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT), one unit per box.November 6, 2013
- FDAEdwards Lifesciences EMBOL-X Glide Protection System; Model Numbers: EXGF24D, EXGF24LLD, EXGF24MMD, EXGF24SSD,November 6, 2013
Questions about this recall
Was DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas. recalled?
Yes. FDA published recall Z-0090-2014 on November 6, 2013. Reported hazard: During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter