McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

Published October 29, 2014 · Recall initiated: June 25, 2014

Summary

FDA records medical device Class II recall Z-0093-2015, published October 29, 2014 and initiated June 25, 2014. Recalling firm: Mckesson Information Solutions LLC; Status: Terminated; Reported quantity: 625 devices; Lot: McKesson Cardiology TM Hemo versions: 11.0HF5, 11.1, 11.1SP5, 11.1SP6, 11.1SP6HF1, 11.1SP6HF3, 12SP1, 12.1, 12.1HF1, 12.1HF2, 12.2, 12.2HF1, 12.2HF2, 12.2HF3, 12.2HF4. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Mckesson Information Solutions LLC
Distributor
Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.
Lot
McKesson Cardiology TM Hemo versions: 11.0HF5, 11.1, 11.1SP5, 11.1SP6, 11.1SP6HF1, 11.1SP6HF3, 12SP1, 12.1, 12.1HF1, 12.1HF2, 12.2, 12.2HF1, 12.2HF2, 12.2HF3, 12.2HF4.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment. recalled?

Yes. FDA published recall Z-0093-2015 on October 29, 2014. Reported hazard: McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i

What should I do if I have this product?

Voluntary: Firm initiated · Letter