McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.
Published October 29, 2014 · Recall initiated: June 25, 2014
Summary
FDA records medical device Class II recall Z-0093-2015, published October 29, 2014 and initiated June 25, 2014. Recalling firm: Mckesson Information Solutions LLC; Status: Terminated; Reported quantity: 625 devices; Lot: McKesson Cardiology TM Hemo versions: 11.0HF5, 11.1, 11.1SP5, 11.1SP6, 11.1SP6HF1, 11.1SP6HF3, 12SP1, 12.1, 12.1HF1, 12.1HF2, 12.2, 12.2HF1, 12.2HF2, 12.2HF3, 12.2HF4. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Mckesson Information Solutions LLC
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.
- Lot
- McKesson Cardiology TM Hemo versions: 11.0HF5, 11.1, 11.1SP5, 11.1SP6, 11.1SP6HF1, 11.1SP6HF3, 12SP1, 12.1, 12.1HF1, 12.1HF2, 12.2, 12.2HF1, 12.2HF2, 12.2HF3, 12.2HF4.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715October 29, 2014
- FDAContinuum Acetabular System Trabecular Metal Shell with Multi Holes, Porous 68mm, Prosthesis, Hip,October 29, 2014
- FDAMcKesson Cardiology" Hemo intended for complete physiological/hemodynamic monitoring, clinical dataOctober 15, 2014
- FDAHorizon Medical Imaging, Picture Archive Communication System (PACS).February 19, 2014
- FDAMcKesson Radiology 12.0, Picture Archiving Communication System (PACSA).January 1, 2014
- FDAMcKesson Horizon Medical Imaging (HMI) Radiological Image Processing System, product versions 4.6.1 to 11.6.,October 9, 2013
Questions about this recall
Was McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment. recalled?
Yes. FDA published recall Z-0093-2015 on October 29, 2014. Reported hazard: McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i
What should I do if I have this product?
Voluntary: Firm initiated · Letter