McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.
Published October 29, 2014 · Recall initiated: June 25, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Mckesson Information Solutions LLC
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.
- Lot
- McKesson Cardiology TM Hemo versions: 11.0HF5, 11.1, 11.1SP5, 11.1SP6, 11.1SP6HF1, 11.1SP6HF3, 12SP1, 12.1, 12.1HF1, 12.1HF2, 12.2, 12.2HF1, 12.2HF2, 12.2HF3, 12.2HF4.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States