GYRUS ACMI, REF 3345 -Everest BiCOAG MOLly Forceps 5mm/45cm REF 3345PK - PKS MOLly Forceps 5mm/45cm REF 3640 -Everest BiCOAG Macro Jaw Forceps 5mm/45cm REF 3640PK- PKS Macro Jaw Forceps 5mm/45cm REF 3844- Everest BiCOAG LP Scissors 5mm/45cm REF 910010PK- PKS MOLly Forceps 5mm/45cm Single-use, sterile devices intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general surgical procedures.
Published October 29, 2014 · Recall initiated: August 27, 2014
Summary
FDA records medical device Class II recall Z-0100-2015, published October 29, 2014 and initiated August 27, 2014. Recalling firm: Gyrus Medical, Inc; Status: Terminated; Reported quantity: 5520 US, 40 OUS; Lot: All product manufactured before July 28, 2014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to an anomaly in the packaging process, it is possible that a defective seal could be present that may allow a breach of the package's sterile barrier and may compromise the sterility of the product.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Gyrus Medical, Inc
- Distributor
- Worldwide Distribution - USA including US: AL, AZ, CA, FL, GA, HI, IL, IN, IA, KY, MA,MI,MS, NJ, NY, NC, OH, PA, SC, TN, UT, and WA. Internationally to ITALY, SPAIN, UNITED KINGDOM, and CHINA.
- Lot
- All product manufactured before July 28, 2014.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"October 29, 2014
- FDAUltra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable powerOctober 29, 2014
- FDADeRoyal(R) SafeLiner Suction Canister Kit, SEMI-RIGID, 1000cc, REF 71-6518, NON-STERILE, Rx Only, generalOctober 29, 2014
- FDAGyrus ACMI Falope-Ring Dilator, item 000878-4. The Falope-Ring Dilator is a conical plastic device that is aAugust 27, 2014
- FDAGyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mmMarch 19, 2014
- FDAGYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting Forceps 5mm/45 cm, Sterile EO. GYRUS ACMI, REF 3740PK,August 28, 2013
Questions about this recall
Was GYRUS ACMI, REF 3345 -Everest BiCOAG MOLly Forceps 5mm/45cm REF 3345PK - PKS MOLly Forceps 5mm/45cm REF 3640 -Everest BiCOAG Macro Jaw Forceps 5mm/45cm REF 3640PK- PKS Macro Jaw Forceps 5mm/45cm REF 3844- Everest BiCOAG LP Scissors 5mm/45cm REF 910010PK- PKS MOLly Forceps 5mm/45cm Single-use, sterile devices intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general surgical procedures. recalled?
Yes. FDA published recall Z-0100-2015 on October 29, 2014. Reported hazard: Due to an anomaly in the packaging process, it is possible that a defective seal could be present that may allow a breach of the package's sterile barrier and may compromise the sterility of the product.
What should I do if I have this product?
Voluntary: Firm initiated · Letter