EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders.
Published October 31, 2012 · Recall initiated: July 2, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Ethicon, Inc.
- Distributor
- Worldwide Distribution -- USA (nationwide) and the countries of Canada, European Union, Singapore, Israel, Mexico and Brazil.
- Lot
- To include all lots and product codes for EVICEL and Omrix pressure regulators.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States