EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders.
Published October 31, 2012 · Recall initiated: July 2, 2012
Summary
FDA records medical device Class II recall Z-0111-2013, published October 31, 2012 and initiated July 2, 2012. Recalling firm: Ethicon, Inc.; Status: Terminated; Reported quantity: 4800 pressure regulators distributed worldwide; 1500 distributed within the US; Lot: To include all lots and product codes for EVICEL and Omrix pressure regulators. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ethicon, Inc.
- Distributor
- Worldwide Distribution -- USA (nationwide) and the countries of Canada, European Union, Singapore, Israel, Mexico and Brazil.
- Lot
- To include all lots and product codes for EVICEL and Omrix pressure regulators.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACoated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358HNovember 13, 2024
- FDAVICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation,June 26, 2024
- FDAPDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation,June 26, 2024
- FDAMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligationJune 26, 2024
- FDASynchron Systems Synchron Calibrator Multi Calibrator Diskettes, Part Number 442600. The SYNCHRONOctober 31, 2012
- FDAAdvanDx GNR Traffic Light PNA FISH, In Vitro Diagnostic for Culture Identification Kit for EscherichiaOctober 31, 2012
Questions about this recall
Was EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders. recalled?
Yes. FDA published recall Z-0111-2013 on October 31, 2012. Reported hazard: Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend
What should I do if I have this product?
Voluntary: Firm initiated · Letter