VIA 27 Microcatheter, Part No. FG27-154-01, Model No. VIA-27-154-01 The VIA 27 is intended for the introduction of non-liquid interventional devices (such as stents/flow diverters) and infusion of diagnostic (such as contrast media) or therapeutic agents into the neuro, peripheral, and coronary vasculature. The catheter is inserted into the body through a small puncture made in the skin and into a blood vessel. The VIA 27 Microcatheter is contraindicated for use with liquid embolic materials, such as n-butyl 2-cyanoacrylate or ethylene vinyl alcohol & DMSO (dimethyl sulfoxide).
Published November 2, 2016 · Recall initiated: September 23, 2016
Summary
FDA records medical device Class II recall Z-0114-2017, published November 2, 2016 and initiated September 23, 2016. Recalling firm: Sequent Medical Inc; Status: Terminated; Reported quantity: 618 units; Lot: 16031101, 16042901, 16051201, 15102107, 15102701, 15111201, 15112003, 16010403, 16011409, 16012901, 16040801, 16041501, 16052603, 16061701. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Sequent Medical is recalling the VIA 27 Microcatheter because it may not adequately re-sheath the interventional device into the catheter as designed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sequent Medical Inc
- Distributor
- Worldwide Distribution - US including NY, MN, FL, MO, MD, TN, NC, OH, CA, KS, TN, WV and Internationally to Czech Republic, Germany, UK, France, Netherlands, Italy, Spain, Finland, Switzerland, Slovenia, Hungary, Northern Ireland, Turkey, and Denmark,
- Lot
- 16031101, 16042901, 16051201, 15102107, 15102701, 15111201, 15112003, 16010403, 16011409, 16012901, 16040801, 16041501, 16052603, 16061701
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVS II Aliquoter System calculator/data processing module, for clinical use Product Usage: Elecysy 411 analyzerNovember 2, 2016
- FDAThe PAXWIRE Occlusion Balloon System consists of an occlusion catheter, inflation handle and syringes. It isNovember 2, 2016
- FDAcobas p 701 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The URISYSNovember 2, 2016
- FDAex-TEM Assay for ROTEM delta Thromboelastometry System. Art. no.: 503-05 US, ex-tem. Used to monitor theNovember 2, 2016
- FDAURISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a computer-controlledNovember 2, 2016
- FDAda Vinci® Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: TheNovember 2, 2016
Questions about this recall
Was VIA 27 Microcatheter, Part No. FG27-154-01, Model No. VIA-27-154-01 The VIA 27 is intended for the introduction of non-liquid interventional devices (such as stents/flow diverters) and infusion of diagnostic (such as contrast media) or therapeutic agents into the neuro, peripheral, and coronary vasculature. The catheter is inserted into the body through a small puncture made in the skin and into a blood vessel. The VIA 27 Microcatheter is contraindicated for use with liquid embolic materials, such as n-butyl 2-cyanoacrylate or ethylene vinyl alcohol & DMSO (dimethyl sulfoxide). recalled?
Yes. FDA published recall Z-0114-2017 on November 2, 2016. Reported hazard: Sequent Medical is recalling the VIA 27 Microcatheter because it may not adequately re-sheath the interventional device into the catheter as designed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter