The brand name of the device is Damon 3MX Bicuspid Bracket, an orthodontic bracket. Part Number 494-4491, Description -Damon 3MX Bicuspid Bracket, Lot Number - 011265039 Product Usage: Damon 3MX Bicuspid Brackets are intended to be a fixed attachment on a tooth which holds an archwire during orthodontic treatment.
Published October 31, 2012 · Recall initiated: April 24, 2012
Summary
FDA records medical device Class III recall Z-0120-2013, published October 31, 2012 and initiated April 24, 2012. Recalling firm: Ormco/Sybronendo; Status: Terminated; Reported quantity: 148; Lot: Part Number 494-4491, Description -Damon 3MX Bicuspid Bracket, Lot Number - 011265039. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm initiated the recall as a result of a misidentification of the brackets. Damon 3MX Bicuspid Brackets are produced with a purple colored ID dot to identify them as maxillary bicuspid brackets; however the affected lot of product listed below was produced with dark blue colored ID dots, which identified them as maxillary central incisor brackets.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ormco/Sybronendo
- Distributor
- Worldwide Distribution - US Nationwide and the countries of Canada, China, Costa Rica, Ecuador, French Polynesia, India, Kuwait, Malaysia, Philippines, Singapore, South Korea, Thailand and UAE.
- Lot
- Part Number 494-4491, Description -Damon 3MX Bicuspid Bracket, Lot Number - 011265039
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOrmco Damon Q2 Orthodontic Metal Bracket, Part No. 491-9834, L3R Hi Tq Hook, Tq +13 degrees, Rx. ManufacturedJune 17, 2020
- FDASnapLink, Ref Part No. 438-2190, L6R SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CEFebruary 26, 2020
- FDASealapex Xpress, Part No. 33639 The brand name of the device is Sealapex Xpress, a root canal filling resin.September 28, 2016
- FDASpirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: AnApril 9, 2014
- FDAOsseotite Tapered Certain Implant 5 X 10mm Model INT510 Biomet 3i Dental Implants are intended for surgicalOctober 31, 2012
- FDAGE Discovery NM/CT670; GE Discovery NM 630 Product Usage: a medical tool intended for use by appropriatelyOctober 31, 2012
Questions about this recall
Was The brand name of the device is Damon 3MX Bicuspid Bracket, an orthodontic bracket. Part Number 494-4491, Description -Damon 3MX Bicuspid Bracket, Lot Number - 011265039 Product Usage: Damon 3MX Bicuspid Brackets are intended to be a fixed attachment on a tooth which holds an archwire during orthodontic treatment. recalled?
Yes. FDA published recall Z-0120-2013 on October 31, 2012. Reported hazard: The firm initiated the recall as a result of a misidentification of the brackets. Damon 3MX Bicuspid Brackets are produced with a purple colored ID dot to identify them as maxillary bicuspid brackets; however the affected lot of product listed below was produced with dark blue colored ID dots, which identified them as maxillary central incisor brackets.
What should I do if I have this product?
Voluntary: Firm initiated · Letter