IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitro diagnostic use with the IMMULITE 1000 Systems Analyzers for the quantitative measurement measurement of cortisol (hydrocortisone, Compound F) in serum, as an aid in the clinical assessment of adrenal status.
Published October 29, 2014 · Recall initiated: September 19, 2014
Summary
FDA records medical device Class II recall Z-0122-2015, published October 29, 2014 and initiated September 19, 2014. Recalling firm: Siemens Healthcare Diagnostics; Status: Terminated; Reported quantity: US: 772 units, Canada: 43 units; Lot: Lots: 381 (exp. 10/31/2014), 382 (exp. 10/31/2014) and 383 (exp. 11/30/2014). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Customer complaints were received for positive bias reported on the controls used on the IMMULITE /IMMULITE 1000 on the Cortisol Assay reagent lots IMMULITE /IMMULITE 1000 (LKCO1) Lots 381,382 and 383 . Subsequently, the firm confirmed a positive bias on the IMMULITE/IMMULITE 1000 (LKCO1) Lots 381, 382 and 383.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics
- Distributor
- Worldwide Distribution: US (nationwide) and internationally to Canada.
- Lot
- Lots: 381 (exp. 10/31/2014), 382 (exp. 10/31/2014) and 383 (exp. 11/30/2014)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that isOctober 29, 2014
- FDAThe Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple totalOctober 29, 2014
- FDAADVIA Centaur Immunoassay System (including refurbished units), This system is intended for professional useOctober 8, 2014
- FDAADVIA Centaur XP Immunoassay System (including refurbished units), This system is intended for professionalOctober 8, 2014
- FDAADVIA Centaur XP Immunoassay System, Catalog Numbers/Siemens Material Numbers (SMN) 10285219, 10316507,September 24, 2014
- FDAADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/SiemensSeptember 24, 2014
Questions about this recall
Was IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitro diagnostic use with the IMMULITE 1000 Systems Analyzers for the quantitative measurement measurement of cortisol (hydrocortisone, Compound F) in serum, as an aid in the clinical assessment of adrenal status. recalled?
Yes. FDA published recall Z-0122-2015 on October 29, 2014. Reported hazard: Customer complaints were received for positive bias reported on the controls used on the IMMULITE /IMMULITE 1000 on the Cortisol Assay reagent lots IMMULITE /IMMULITE 1000 (LKCO1) Lots 381,382 and 383 . Subsequently, the firm confirmed a positive bias on the IMMULITE/IMMULITE 1000 (LKCO1) Lots 381, 382 and 383.
What should I do if I have this product?
Voluntary: Firm initiated · Letter