The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the Persona Tibial Articular Inserter is used to seat the articular surface implant onto the tibial baseplate.
Published October 29, 2014 · Recall initiated: October 7, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected product did not undergo heat treatment hardening step leading to the potential for deformation of the tip.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) including the states of NY, OR, NC, OH, TX, FL, CO, IN, TN, PA, CA, KS, MO, WI, VA, MN, NJ, UT, MA, GA, MI, OK, AZ, MS, IL and AL., and the countries of Korea, Singapore, Canada and Japan
- Lot
- Part Number 42-5299-001-00 Lots: 56572679; 56573170; 56572918; 56573378; 56572975; 56573379; 56572978; 56574223; 56572981; 56574291; 56572982; 56574292; 56573071; 56574316; 56573115; 56574317; 56573116; 56574318; 56573154.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States