The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the Persona Tibial Articular Inserter is used to seat the articular surface implant onto the tibial baseplate.
Published October 29, 2014 · Recall initiated: October 7, 2014
Summary
FDA records medical device Class II recall Z-0123-2015, published October 29, 2014 and initiated October 7, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 618 units; Lot: Part Number 42-5299-001-00 Lots: 56572679; 56573170; 56572918; 56573378; 56572975; 56573379; 56572978; 56574223; 56572981; 56574291; 56572982; 56574292; 56573071; 56574316; 56573115; 56574317; 56573116; 56574318; 56573154. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected product did not undergo heat treatment hardening step leading to the potential for deformation of the tip.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) including the states of NY, OR, NC, OH, TX, FL, CO, IN, TN, PA, CA, KS, MO, WI, VA, MN, NJ, UT, MA, GA, MI, OK, AZ, MS, IL and AL., and the countries of Korea, Singapore, Canada and Japan
- Lot
- Part Number 42-5299-001-00 Lots: 56572679; 56573170; 56572918; 56573378; 56572975; 56573379; 56572978; 56574223; 56572981; 56574291; 56572982; 56574292; 56573071; 56574316; 56573115; 56574317; 56573116; 56574318; 56573154.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAEmtriva (emtricitabine) Oral Solution, 10 mg/mL, 170 mL bottles, Rx only, Manufactured for Gilead Sciences,November 5, 2014
- FDAIMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitroOctober 29, 2014
Questions about this recall
Was The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the Persona Tibial Articular Inserter is used to seat the articular surface implant onto the tibial baseplate. recalled?
Yes. FDA published recall Z-0123-2015 on October 29, 2014. Reported hazard: Affected product did not undergo heat treatment hardening step leading to the potential for deformation of the tip.
What should I do if I have this product?
Voluntary: Firm initiated · Letter