Back to search
FDAFDA · Food and drugs·Recall number Z-0127-2014

Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.

Published November 13, 2013 · Recall initiated: November 4, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken the device and may result in a fracture of the auger and/or cannula. This could result in the need for additional medical intervention to remove the broken piece, pain or loss of m

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Instruments Div. of Stryker Corporation
Distributor
Nationwide Distribution including the states of AR, AZ, CA, CO, FL, GA, IL, KS, MS, ND, and NE.
Lot
Part Number(s) 0407-251-000, 0407-260-000, 0407-265-000, 0407-266-000, 0407-280-000, and 0407-281-000.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card