Back to search
FDAFDA · Food and drugs·Recall number Z-0131-2023

LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.

Published November 2, 2022 · Recall initiated: August 5, 2022

Summary

FDA records medical device Class II recall Z-0131-2023, published November 2, 2022 and initiated August 5, 2022. Recalling firm: Datascope Corporation; Status: Ongoing; Reported quantity: 4414 OUS; Lot: UDI-DI: 10607567106571 Lot codes: 3000210759 3000210758 3000203002 3000198455 3000144227 3000142368 3000141069 3000140532 3000134522 3000130423 3000126591. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy. Therapy delays may impact those patients who are more hemodynamically unstable.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Datascope Corporation
Distributor
No US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Columbia, Ecuador, Egypt, France, Indonesia, Israel, Italy, Jordan, Kenya, Malaysia, Mexico, Netherlands, Pakistan, Poland, Portugal, South Korea, Russia, Spain, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay.
Lot
UDI-DI: 10607567106571 Lot codes: 3000210759 3000210758 3000203002 3000198455 3000144227 3000142368 3000141069 3000140532 3000134522 3000130423 3000126591
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter. recalled?

Yes. FDA published recall Z-0131-2023 on November 2, 2022. Reported hazard: The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy. Therapy delays may impact those patients who are more hemodynamically unstable.