da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)

Published November 2, 2022 · Recall initiated: September 20, 2022

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
Lot
Software version P10 (OS4 v10.0.0/A70_P10_B738) SOFTWARE,EMBEDDED RLS,IS4000 AND IS4200,A70_P10_B738 (Model: 610092-738) on Model Name/Model Number/UDI-DI: ASSY,PSCART,IS4000,4-ARM,P10/380652-55/00886874110720, ASSY,SSC,IS4000,P10/380677-28/00886874110744, ASSY,VSS VISION SYSTEM,IS4000,P10/381121-45/00886874110898, ASSY,PSCART,IS4200,P10/380620-47/00886874115404, ASSY,VSC,IS4000,RECONDITIONED,P10/380721-15/00886874115343, ASSY,SSC,IS4000,RECONDITIONED,P10/380723-16/00886874115374
Risk classification
Class II
Area (FDA)
device
Country
United States