Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with Duraflo Coating, DIITFA02225, Sterile, RX only For Single Use Only, Product Usage: Edwards Lifesciences femoral access cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Femoral access cannulae may be used in pediatric populations or adult populations based on flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Published November 13, 2013 · Recall initiated: August 26, 2013
Summary
FDA records medical device Class II recall Z-0134-2014, published November 13, 2013 and initiated August 26, 2013. Recalling firm: Edwards Lifesciences, LLC; Status: Terminated; Reported quantity: 80 units; Lot: Lot number 59388992 expiry 1/1/16. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Edwards Lifesciences, LLC
- Distributor
- Nationwide Distribution.
- Lot
- Lot number 59388992 expiry 1/1/16
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGeneric Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery SystemNovember 11, 2026
- FDAEdwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EOMay 20, 2026
- FDABrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS SoftwareJuly 16, 2025
- FDABrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16July 16, 2025
- FDAThe EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power perNovember 13, 2013
- FDAEdwards Lifesciences Dispersion Aortic Perfusion Cannula with Duraflo Coating, Sterile, RX only For Single UseNovember 13, 2013
Questions about this recall
Was Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with Duraflo Coating, DIITFA02225, Sterile, RX only For Single Use Only, Product Usage: Edwards Lifesciences femoral access cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Femoral access cannulae may be used in pediatric populations or adult populations based on flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired. recalled?
Yes. FDA published recall Z-0134-2014 on November 13, 2013. Reported hazard: Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.
What should I do if I have this product?
Voluntary: Firm initiated · Letter