FDAFDA · Food and drugs·Recall number Z-0136-2015

Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Published November 5, 2014 · Recall initiated: June 10, 2014

Summary

FDA records medical device Class II recall Z-0136-2015, published November 5, 2014 and initiated June 10, 2014. Recalling firm: Animas Corporation; Status: Terminated; Reported quantity: 1028; Lot: Model Number(s): 11220008 PMPW/OKIT,VIBE,BLU,mg,IL 11220108 PMPW/OKIT,VIBE,BLK,mg,IL 11220208 PMPW/OKIT,VIBE,SIL,mg,IL 11220408 PMPW/OKIT,VIBE,PGL,mg,IL 11220508 PMPW/OKIT,VIBE,LGN,mg,IL 11251008 REV KIT,VIBE,BLU,mg,IL 11251108 REV KIT,VIBE,BLK,mg,IL 11251208 REV KIT,VIBE,SIL,mg,IL 11251408 REV KIT,VIBE,PGL,mg,IL 11251508 REV KIT,VIBE,LGN,mg,IL 11451008 REV KIT,VIBEPL,BLU,mg,IL 11451108 REV KIT,VIBEPL,BLK,mg,IL 11451208 REV KIT,VIBEPL,SIL,mg,IL 11451408 REV KIT,VIBEPL,PGL,mg, IL 11451508 REV KIT,VIBEPL,LGN,mg,IL 11420008 PMPW/OKIT,VIBEPL,BLU,mg,IL 11420108 PMPW/OKIT,VIBEPL,BLK,mg,IL 11420208 PMPW/OKIT,VIBEPL,SIL,mg,IL 11420408 PMPW/OKIT,VIBEPL,PGL,mg,IL 11420508 PMPW/OKIT,VIBEPL,LGN,mg,IL. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Animas determined that the Hebrew language translation in the Vibe Insulin Infusion Pump was incorrect and could lead to potential patient confusion. In addition, one line of Hebrew text was omitted from the display screen.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Animas Corporation
Distributor
International Distribution in Israel only.
Lot
Model Number(s): 11220008 PMPW/OKIT,VIBE,BLU,mg,IL 11220108 PMPW/OKIT,VIBE,BLK,mg,IL 11220208 PMPW/OKIT,VIBE,SIL,mg,IL 11220408 PMPW/OKIT,VIBE,PGL,mg,IL 11220508 PMPW/OKIT,VIBE,LGN,mg,IL 11251008 REV KIT,VIBE,BLU,mg,IL 11251108 REV KIT,VIBE,BLK,mg,IL 11251208 REV KIT,VIBE,SIL,mg,IL 11251408 REV KIT,VIBE,PGL,mg,IL 11251508 REV KIT,VIBE,LGN,mg,IL 11451008 REV KIT,VIBEPL,BLU,mg,IL 11451108 REV KIT,VIBEPL,BLK,mg,IL 11451208 REV KIT,VIBEPL,SIL,mg,IL 11451408 REV KIT,VIBEPL,PGL,mg, IL 11451508 REV KIT,VIBEPL,LGN,mg,IL 11420008 PMPW/OKIT,VIBEPL,BLU,mg,IL 11420108 PMPW/OKIT,VIBEPL,BLK,mg,IL 11420208 PMPW/OKIT,VIBEPL,SIL,mg,IL 11420408 PMPW/OKIT,VIBEPL,PGL,mg,IL 11420508 PMPW/OKIT,VIBEPL,LGN,mg,IL
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature. recalled?

Yes. FDA published recall Z-0136-2015 on November 5, 2014. Reported hazard: Animas determined that the Hebrew language translation in the Vibe Insulin Infusion Pump was incorrect and could lead to potential patient confusion. In addition, one line of Hebrew text was omitted from the display screen.

What should I do if I have this product?

Voluntary: Firm initiated · Letter