FDAFDA · Food and drugs·Recall number Z-0136-2025

bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option.

Published October 30, 2024 · Recall initiated: September 4, 2024

Summary

FDA records medical device Class II recall Z-0136-2025, published October 30, 2024 and initiated September 4, 2024. Recalling firm: B-K Medical A/S; Status: Ongoing; Reported quantity: 1648 devices; Lot: All serial numbers, GTIN 05704916000264. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
B-K Medical A/S
Distributor
Worldwide - US Nationwide - There was also government/military distribution and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Lot
All serial numbers, GTIN 05704916000264.
Risk classification
Class II
Area (FDA)
device
Country
Denmark

Related recalls

Questions about this recall

Was bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option. recalled?

Yes. FDA published recall Z-0136-2025 on October 30, 2024. Reported hazard: The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

What should I do if I have this product?

Voluntary: Firm initiated · Letter