bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option.
Published October 30, 2024 · Recall initiated: September 4, 2024
Summary
FDA records medical device Class II recall Z-0136-2025, published October 30, 2024 and initiated September 4, 2024. Recalling firm: B-K Medical A/S; Status: Ongoing; Reported quantity: 1648 devices; Lot: All serial numbers, GTIN 05704916000264. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B-K Medical A/S
- Distributor
- Worldwide - US Nationwide - There was also government/military distribution and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
- Lot
- All serial numbers, GTIN 05704916000264.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Denmark
Related recalls
- FDASoltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation,October 30, 2024
- FDASynchroMed II Infusion Pump. Product Number: 8637-20October 30, 2024
- FDASoltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation,October 30, 2024
- FDAQuantimetrix Urine Chemistry Control Set Dipper Urinalysis Dipstick Testing Levels 1 & 2 6 X 15 mL-IVDOctober 30, 2024
- FDA(1) Centurion Sterile CirClamp, Reorder #200CR; (2) Centurion CirClamp with 1.1cm Bell and Insert, ReorderOctober 30, 2024
- FDAbkActiv Ultrasound System 2300, Model numbers 2300-56 and 2300-66.January 31, 2024
Questions about this recall
Was bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option. recalled?
Yes. FDA published recall Z-0136-2025 on October 30, 2024. Reported hazard: The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.
What should I do if I have this product?
Voluntary: Firm initiated · Letter