Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV.
Published November 5, 2014 · Recall initiated: September 29, 2014
Summary
FDA records medical device Class II recall Z-0143-2015, published November 5, 2014 and initiated September 29, 2014. Recalling firm: Merge Healthcare, Inc.; Status: Terminated; Reported quantity: 52; Lot: Merge Hemo 9.x with Nellcor SpO2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subject to displaying a stale value.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Merge Healthcare, Inc.
- Distributor
- Nationwide Distribution including the states of AZ, CA, FL, IL, IA, MA, MO, OH, SC, TN, and WI.
- Lot
- Merge Hemo 9.x with Nellcor SpO2
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMerge PACS and Merge OrthoPACS, medical imaging diagnostic workstationSeptember 2, 2026
- FDAMerge OrthoPACS, medical imaging diagnostic workstationAugust 12, 2026
- FDAMerge PACS, medical imaging diagnostic workstationAugust 12, 2026
- FDAVERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0,March 26, 2025
- FDADMI Duro-Med Industries Rollator Transport Chair; Wheel Size: 7 1/4"; Overall Width: 23 1/2"; Seat Height: 21November 5, 2014
- FDAIMRIS Operating Room table (ORT200 and ORT300). Intended for use during diagnostic examinations or surgicalNovember 5, 2014
Questions about this recall
Was Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV. recalled?
Yes. FDA published recall Z-0143-2015 on November 5, 2014. Reported hazard: It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subject to displaying a stale value.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail