COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological
Published November 12, 2014 · Recall initiated: September 25, 2014
Summary
FDA records medical device Class II recall Z-0168-2015, published November 12, 2014 and initiated September 25, 2014. Recalling firm: Iba Dosimetry Gmbh; Status: Terminated; Reported quantity: 124 units. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Iba Dosimetry Gmbh
- Distributor
- Worldwide Distribution-Distributed in the US including Guam and the countries of Germany, Austria, France, UK, Poland, Finland, The Netherlands, Spain, Turkey, Russia, and South Africa, Algeria, Mexico, Brazil, Venezuela, Columbia, Panama, China, Japan Philippines, Korea, India, Singapore, Thailand, Taiwan, Hong Kong, and Malaysia.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAPrescription eyeglass safety lenses. Vision correctionNovember 12, 2014
- FDAElite XPL (Pulsed Light) Handpiece for the Cynosure Elite MPX Laser SystemNovember 12, 2014
- FDACADD-Solis Medication Safety Software, Administrator CD, Version 3.1, REF 21-2194-0301-01. Product Usage: TheNovember 12, 2014
- FDAStryker SmartLife Large Aseptic Housing REF 7126-120-000 Rx Only The SmartLife" Aseptic Housings are designedNovember 12, 2014
- FDAUNIVERSAL Modular Electric/Battery Handpiece for surgery The electric handpiece is packaged within a foam bagNovember 12, 2014
- FDAiba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiologicalDecember 18, 2013
Questions about this recall
Was COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological recalled?
Yes. FDA published recall Z-0168-2015 on November 12, 2014. Reported hazard: Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit