Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
Published November 8, 2023 · Recall initiated: September 15, 2023
Summary
FDA records medical device Class I recall Z-0170-2024, published November 8, 2023 and initiated September 15, 2023. Recalling firm: Asensus Surgical Inc; Status: Terminated; Reported quantity: 21 units; Lot: UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Asensus Surgical Inc
- Distributor
- US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania
- Lot
- UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANature's Path Organic Dark Chocolate Chip Waffles, 7 oz., UPC 0 5844916701 3November 8, 2023
- FDA***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgicalNovember 8, 2023
- FDANature's Path Organic Pumpkin Spice Waffles, 7 oz., UPC 0 5844959077 4November 8, 2023
- FDACoapt Dome Electrode, Gen2 system, cutaneous electrodeNovember 8, 2023
- FDAJayRobb Vanilla Flavored Egg White Protein Powder. Net Weight 24 oz (1.5 lbs) 680g Dietary Supplement. UPC 6November 8, 2023
- FDAUnity Total Knee System, Model Number 112.001.34. Primary and revision total knee arthroplastyNovember 8, 2023
Questions about this recall
Was Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system recalled?
Yes. FDA published recall Z-0170-2024 on November 8, 2023. Reported hazard: Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter