FDAFDA · Food and drugs·Recall number Z-0176-2022

ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi- site laparoscopic surgery during power morcellation and removal.

Published November 3, 2021 · Recall initiated: September 21, 2021

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, which will make tissue encapsulation and bag closure more difficult, introducing the risk of trapping the small bowel/viscera in the bag at closure resulting in patient injury.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
US Nationwide distribution to states of: AL, CA, CT, FL, IL, MD, MI, MN, MO, NE, NY, OH, OK, OR, PA, SC, UT, VA, WA, WI, and WV ; and International (foreign) distribution to: Europe.
Lot
Lots 647572 and 667060
Risk classification
Class II
Area (FDA)
device
Country
United States