FDAFDA · Food and drugs·Recall number Z-0180-2014
CARESTREAM DRX -Mobile Retrofit Kit, Catalog number: 1019397. Made in U.S.A. by Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 to perform radiographic x-ray examinations on all pediatric and adult patients
Published November 13, 2013 · Recall initiated: August 29, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Carestream Health Inc. has conducted a voluntary field action which includes the installation of a software patch on the following: DRX-Revolution Mobile X-Ray, DRX-Mobile and DRX Transportable Systems with DirectView Version 5.5 and 5.6 console software when used with DRX-1 Detector version 69 firmware. Carestream recently received reports of a non-diagnostic image on exams when a Normal Exposure
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Carestream Health Inc.
- Distributor
- Nationwide distribution:AL, AZ, AR, CA, CO, CT, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MI, MO, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VT, VA, WA, WI, and WV.
- Lot
- Model: DRX Mobile Serial numbers: all serial numbers manufactured with version 5.5 software and version 69 firmware
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?