Back to search
FDAFDA · Food and drugs·Recall number Z-0184-2022

FIVE S 3.5x65, Flexible Intubation Videoscope, REF: 091361-01 (1 Unit), and 091361-06 (6 Units)

Published November 10, 2021 · Recall initiated: September 10, 2021

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There is a risk that the patient may be exposed to a higher risk of infection.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Karl Storz Endoscopy
Distributor
US: SC, NH, UT, LA, FL, TX, MA, MO, VA, IA, CA, IL, NC, VT, AL, GA, PA, TN, WI, NY, CO, AZ, NV, ME, MS, DC, IN, KY, MD, OH, MI, WA, KS, OR, AR, MN, NJ, NM, WV, CT. OUS: Canada
Lot
All LOTs with remaining shelf life
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.