CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.
Published October 31, 2018 · Recall initiated: August 28, 2018
Summary
FDA records medical device Class I recall Z-0185-2019, published October 31, 2018 and initiated August 28, 2018. Recalling firm: Alcon Research, LTD.; Status: Terminated; Reported quantity: 30561 units; Lot: GTIN: 00380657540037 All lot/serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alcon Research, LTD.
- Distributor
- Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.
- Lot
- GTIN: 00380657540037 All lot/serial numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDALegion FMS PakAugust 23, 2023
- FDAFIRE-ROASTED ORGANIC SUPER SWEET CORN; Sold in 1500 lb. containers. Brand Name: Jon-Lin®October 31, 2018
- FDAAPTIO RACK INPUT MODULEOctober 31, 2018
- FDAFire-Roasted 1/2" Diced Red Onions; Sold in 25 lb. containers. Brand Name: Jon-Lin®October 31, 2018
- FDAAlcon Custom PakNovember 8, 2017
- FDAThis voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocarOctober 3, 2012
Questions about this recall
Was CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use. recalled?
Yes. FDA published recall Z-0185-2019 on October 31, 2018. Reported hazard: Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
What should I do if I have this product?
Voluntary: Firm initiated · Letter