FDAFDA · Food and drugs·Recall number Z-0192-2025
medex TranStar 84in (213cm) Single Monitoring Kit 10/EA, Product Code REF MX9505T
Published October 30, 2024 · Recall initiated: September 24, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
What to do
Voluntary: Firm initiated
Recall details
- Manufacturer
- Smiths Medical ASD, Inc.
- Distributor
- Worldwide distribution.
- Lot
- UDI/DI 50351688503990 (case), 10351688503992 (pouch), Lot Numbers: 4277670, 4287645, 4287649, 4299776, 4277696, 4308930, 4299811, 4308934, 4299812, 4323148, 4264943, 4264939, 4277694, 4264950, 4299779, 4308962, 4323175, 4328515, 4299788, 4299789, 4299790, 4299801, 4299802, 4308954, 4308959, 4308963, 4321917, 4308947, 4334499, 4299786, 4355732, 4355735, 4346115, 4299780, 4355730, 4355727, 4334501, 4346096, 4346097, 4346124, 4355733, 4346109, 4355728, 4355740, 4355750, 4355749, 4355746, 4346120, 4355757, 4355753
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States