Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.

Published November 20, 2013 · Recall initiated: October 28, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Toshiba American Medical Systems Inc
Distributor
USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.
Lot
Lot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A5C1222010, A5C1232011, A5C1252012, A5C1262013, A5C1282014, A5C1292015, A5C12X2016, A5C12Y2017, A5C12Z2018, A5D1312019, A5D1322020, A5D1362021, XIB-A5A1082001, XIB-A5B11Y2006, A5D1372022, A5B1182002.
Risk classification
Class II
Area (FDA)
device
Country
United States