C3 Wave App, v. 2.0.5

Published November 6, 2019 · Recall initiated: October 1, 2018

Summary

FDA records medical device Class II recall Z-0222-2020, published November 6, 2019 and initiated October 1, 2018. Recalling firm: Medical Components, Inc dba MedComp; Status: Terminated; Reported quantity: 24; Lot: v. 2.0.5; all units. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When the iPad is updated with Apple iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medical Components, Inc dba MedComp
Distributor
Distributed to accounts in AZ, CA, CO, FL, IL, LA, MA, ME, NC, OH, TN, VA, and WV. Foreign distribution to Canada, Croatia, Czech Republic, Great Britain, Greece, Italy, Netherlands, Slovenia, and Sweden.
Lot
v. 2.0.5; all units
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was C3 Wave App, v. 2.0.5 recalled?

Yes. FDA published recall Z-0222-2020 on November 6, 2019. Reported hazard: When the iPad is updated with Apple iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit