FDAFDA · Food and drugs·Recall number Z-0223-2025

A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the monitor), (2) a plastic cartridge enclosing dry reagent strips, and (3) a sample dilution kit for: collecting the blood sample, mixing the sample with the required pre-treatment solution, and delivering the sample to the cartridge.

Published October 30, 2024 · Recall initiated: September 18, 2024

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may result in inaccurate test results as components within each lot number are meant to be used together, and not mixed between lots. Therefore, the test results may possibly lead to Hemoglobin A1C positive or negative bias results. While the product is for monitoring only, over recovery with a higher than accurate value can be an indication for increased therapy; such as physical activity, weight reduction, meal management, pharmaceutical therapy.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Japan, South Africa.
Lot
UDI-DI: 00381933021011 Lot Numbers: 9331803, 2323019
Risk classification
Class II
Area (FDA)
device
Country
United States