Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is indicated for use in retinal photocoagulation for treatment of ocular fundus diseases.
Published November 26, 2014 · Recall initiated: October 24, 2012
Summary
FDA records medical device Class II recall Z-0224-2015, published November 26, 2014 and initiated October 24, 2012. Recalling firm: Nidek Inc; Status: Terminated; Reported quantity: 21affected devices; Lot: Model MC-500 Serial numbers: 50277 50310 50330 50331 50357 50358 50383 50408 50417 50418 50425 50426 50439 50440 50445 50446 50471 50514 50311 50359 50382,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.
What to do
Voluntary: Firm initiated · Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Nidek Inc
- Distributor
- US Distribution in the states of: NJ, SC, IL, MI, NY, CA, TX, OH, IN, UT, and MN.
- Lot
- Model MC-500 Serial numbers: 50277 50310 50330 50331 50357 50358 50383 50408 50417 50418 50425 50426 50439 50440 50445 50446 50471 50514 50311 50359 50382,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASPECULAR MICROSCOPE CEM-530; Software version 1.08 and 1.09. Opthalmic: The NIDEK Specular Microscope CEM-530September 7, 2016
- FDAMC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic device used for retinalNovember 26, 2014
- FDANidek Excimer Laser Corneal Surgery System, model EC-5000 CXIII. Opthalmic Device. The system is composed as aNovember 26, 2014
- FDAThe MectaLIF Oblique Handle, model number 03.22.10.0262, is used to insert the MectaLIF oblique spinal cageNovember 26, 2014
- FDASideFire Directional Laser Probe, REF: NBP101-01, NBP201-01, NBP301-01, NBP401-01, NBP501-01. These laserNovember 26, 2014
- FDAGYC-1000 Green Laser Photocoagulator system The Nidek Green Laser Photocoagulator Model GYC -1000 is indicatedAugust 20, 2014
Questions about this recall
Was Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is indicated for use in retinal photocoagulation for treatment of ocular fundus diseases. recalled?
Yes. FDA published recall Z-0224-2015 on November 26, 2014. Reported hazard: When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.
What should I do if I have this product?
Voluntary: Firm initiated · Visit