BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack.
Published November 24, 2021 · Recall initiated: September 17, 2021
Summary
FDA records medical device Class II recall Z-0236-2022, published November 24, 2021 and initiated September 17, 2021. Recalling firm: Bio-Rad Laboratories, Inc.; Status: Ongoing; Reported quantity: 791 units; Lot: Catalog Number: 12000651 UDI Code: 03610520652770 Lot Number: 52953. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Bio-Rad Laboratories, Inc.
- Distributor
- Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WV; O.U.S.(foreign) countries of: France, Germany, Israel, Italy, Russian Federation, South Africa, Spain, and Taiwan.
- Lot
- Catalog Number: 12000651 UDI Code: 03610520652770 Lot Number: 52953
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASMART FLEX 6x150 BIL, 120cm The S.M.A.R.T.¿ Flex Biliary Stent System is indicated for use in the palliationNovember 24, 2021
- FDAREGARD Vitrectomy TrayNovember 24, 2021
- FDASimply Fresh Salads - Cranberry Walnut Salad with Very Berry Vinaigrette; sweet-tart cranberries & Walnuts,November 24, 2021
- FDAVariant II beta-thalassemia Short Program Model Numbers 270-2103 (250 tests) and 270-2154 (500 tests), Bio-RadMay 14, 2014
- FDAD-10" Hemoglobin Testing System, Model number: 220-0220 Manufactured by: Bio-Rad Laboratories, Inc. 4000April 2, 2014
- FDAD-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories,February 19, 2014
Questions about this recall
Was BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack. recalled?
Yes. FDA published recall Z-0236-2022 on November 24, 2021. Reported hazard: Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter