Rosebud Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and Temperature (TEMP) on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians

Published November 6, 2019 · Recall initiated: October 1, 2019

Summary

FDA records medical device Class II recall Z-0238-2020, published November 6, 2019 and initiated October 1, 2019. Recalling firm: Mindray DS USA, Inc. dba Mindray North America; Status: Terminated; Reported quantity: 9 units; Lot: NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Mindray DS USA, Inc. dba Mindray North America
Distributor
US Nationwide distributions.
Lot
NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Rosebud Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and Temperature (TEMP) on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians recalled?

Yes. FDA published recall Z-0238-2020 on November 6, 2019. Reported hazard: NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function

What should I do if I have this product?

Voluntary: Firm initiated · Letter