Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #063-7591-7 The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-5000 performs analyses using the following methods: RF/DC Detection Method, Sheath Row DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. Using the same reagents as the XE-2100, the XE-5000 automatically classifies cells from blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.

Published November 14, 2012 · Recall initiated: June 1, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Sysmex America, Inc.
Distributor
Nationwide distribution: USA including Puerto Rico.
Lot
catalog #063-7591-7, all serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States